Specifications

<< Click to Display Table of Contents >>

Navigation:  Instruments > M.O.L.E. EV6 >

Specifications

M.O.L.E.® EV6 Thermal Profiler

INPUTS:

ECD Micro-Thermocouples: Type K, T, J, E, N, R, S & B (Type K Special Limits of error)

%RH: ECD VaporWATCH RH Sensor

PHYSICAL DIMENSIONS (L x W x H):

152.4mm (6”) x 89mm (3.5”) x 9.41mm (0.37”)

DISPLAY:

PS TFT Display

240 x 400 Dot Matrix

High Brightness LED Backlight

Any Angle Viewing Direction

COMMUNICATIONS:

USB 2.0 Compliant (charge; upload; download)

Cable - USB Type A to USB-C, 1m (3’)

TEMPERATURE MEASUREMENT RANGE:

Measurement range: -200°C to 1,200°C (-328°F to 2,192°F)

INTERNAL OPERATING TEMPERATURE RANGE:

-40°C to 80°C (-40°F to 176°F)

ABSOLUTE MAXIMUM WARRANTEED INTERNAL TEMPERATURE:

87°C (188.6°F)* M.O.L.E.® EV6 Profiler automatically stops operating when the internal temperature exceeds 80°C (176°F)

Warning

*WARNING: Exceeding this temperature may permanently damage the equipment!

ACCURACY:

+/- 1°C (1.8°F). Channel to channel compliance within 1°C.

RESOLUTION:

0.0056°C (0.01°F)*

NOISE REJECTION:

50-60 hrz*

PROGRAMMABLE SAMPLING INTERVAL:

100 samples/second to 10 minutes/sample

HARDWARE REAL-TIME CLOCK/CALENDAR:

±1 minute/month at 25°C

OPEN THERMOCOUPLE DETECTION.

All 6 channels

DATA POINTS:

16 million (2.6 million at 6 channels) or 200 data runs (whichever comes first)

 

Internal Battery:

BATTERY:

LiPo (Lithium Polymer ion) – Rechargeable, 1100mAh

CHARGE PROTECTION:

Optimized for lithium polymer

CHARGE RUNTIME:

Full – 6 to 8 Hours

CHARGE TIME:

<4 hours

EXPECTED POWER PACK LIFE:

300-400 charging cycles or 3 years (minimum).

CHARGER INPUT CURRENT:

60 ma via USB Interface

OPERATING RANGE:

4.2V to 3.2V

 

Internal Battery Charger:

INPUT:

Voltage Rating: 100 to 240 VAC

Current: 200mA(RMS) max. @ 115VAC

80mA(RMS) max. @ 230VAC

Frequency: 50/60Hz

OUTPUT:

5V DC, Load 0A MIN. 1A MAX.

MAINS VOLTAGE FLUCTUATIONS:

+ 10%

Thermocouple Specifications:

Thermocouples: Type K, Micro-Connector, Glass insulated. (Other insulation types available)

TIME CONSTANT*

Wire Size

mm (inches)

Still Air

427°C/38°C

800°F/100°F

60 Ft./Sec. Air

427°C/38°C

800/100°F

Still H2O

93°C/38°C

200°F/100°F

0.025 (0.001)

0.125 (0.005)

0.381 (0.015)

0.05 sec.

1.0 sec.

10.0 sec.

0.004 sec.

0.08 sec.

0.80 sec.

0.002 sec.

0.04 sec.

0.40 sec.

* The time constant is defined as the time required to reach 63.2% of an instantaneous temperature change.

 

Environmental Limitation Specifications:

M.O.L.E.® EV6 Profiler:

Maximum relative humidity 80% RH from 0°C to 31°C, decreasing linearly to 17% RH at 50°C.

Pollution Degree 2 (Normally only dry pollution, but with temporary conductivity caused by condensation)

Battery Charger:

Temperature: (Operation) 0 to +45°C (32 to 113°F)

Temperature: (Non-Operation) -20 to +75°C (-4 to 167°F)

Humidity: (Operation) 20 to 90%

NOTE:Tests have shown that because of the sensitive nature of the measurement and logic circuits, the following precautions must be observed:

Minimize exposure to ESD Events. If the M.O.L.E. Profiler or one of the thermocouples receives an 8kV electrostatic discharge during the data collection sequence, the M.O.L.E. Profiler may switch itself “OFF”. To retrieve the data, simply upload the data that was recorded prior to the event. If a new data collection sequence is started without uploading first, the existing data stored in the M.O.L.E. Profiler will be lost.

Keep strong electromagnetic fields away. The thermocouple wires serve as an antenna for electromagnetic radiation. If field strength of 3 volts per meter is present (usually due to close proximity of radio transmitters) while the M.O.L.E. Profiler is collecting data, the accuracy of the data may be compromised. This interference is compounded by using long thermocouple wires, as well as the effect of the thermocouple wires acting as a “tuned” antenna.