M.O.L.E.® EV20 Thermal Profiler
INPUTS: |
•Standard Mini Thermocouples: Type K, T, J, E, N, R, S & B (Type K Special Limits of error) •%RH: ECD VaporWATCH RH Sensor |
|
PHYSICAL DIMENSIONS (L x W x H): |
410mm (16.14”) x 89mm (3.5”) x 9.41mm (0.37”) |
|
DISPLAY: |
•PS TFT Display •240 x 400 Dot Matrix •High Brightness LED Backlight •Any Angle Viewing Direction |
|
COMMUNICATIONS: |
•USB 2.0 Compliant (charge; upload; download) •Cable - USB Type A to USB-C, 1m (3’) |
|
TEMPERATURE MEASUREMENT RANGE: |
Measurement range: -200°C to 1,200°C (-328°F to 2,192°F) |
|
INTERNAL OPERATING TEMPERATURE RANGE: |
-40°C to 80°C (-40°F to 176°F) |
|
ABSOLUTE MAXIMUM WARRANTEED INTERNAL TEMPERATURE: |
87°C (188.6°F)* M.O.L.E.® EV20 Profiler automatically stops operating when the internal temperature exceeds 80°C (176°F) |
|
*WARNING: Exceeding this temperature may permanently damage the equipment! |
||
ACCURACY: |
+/- 1°C (1.8°F). Channel to channel compliance within 1°C. |
|
RESOLUTION: |
0.0056°C (0.01°F)* |
|
NOISE REJECTION: |
50-60 hrz* |
|
PROGRAMMABLE SAMPLING INTERVAL: |
20 samples/second to 10 minutes/sample |
|
HARDWARE REAL-TIME CLOCK/CALENDAR: |
±1 minute/month at 25°C |
|
OPEN THERMOCOUPLE DETECTION. |
All 20 channels |
|
DATA POINTS: |
16 million (2.6 million at 20 channels) or 200 data runs (whichever comes first) |
Internal Battery:
BATTERY: |
LiPo (Lithium Polymer ion) – Rechargeable, 1100mAh |
CHARGE PROTECTION: |
Optimized for lithium polymer |
CHARGE RUNTIME: |
Full – 6 to 8 Hours |
CHARGE TIME: |
<4 hours |
EXPECTED POWER PACK LIFE: |
300-400 charging cycles or 3 years (minimum). |
CHARGER INPUT CURRENT: |
60 ma via USB Interface |
OPERATING RANGE: |
4.2V to 3.2V |
Internal Battery Charger:
INPUT: |
Voltage Rating: 100 to 240 VAC Current: 200mA(RMS) max. @ 115VAC 80mA(RMS) max. @ 230VAC Frequency: 50/60Hz |
OUTPUT: |
5V DC, Load 0A MIN. 1A MAX. |
MAINS VOLTAGE FLUCTUATIONS: |
+ 10% |
Thermocouple Specifications:
Thermocouples: Type K, Mini-Connector, Glass insulated. (Other insulation types available) TIME CONSTANT* |
|||
Wire Size mm (inches) |
Still Air 427°C/38°C 800°F/100°F |
60 Ft./Sec. Air 427°C/38°C 800/100°F |
Still H2O 93°C/38°C 200°F/100°F |
0.025 (0.001) 0.125 (0.005) 0.381 (0.015) |
0.05 sec. 1.0 sec. 10.0 sec. |
0.004 sec. 0.08 sec. 0.80 sec. |
0.002 sec. 0.04 sec. 0.40 sec. |
* The time constant is defined as the time required to reach 63.2% of an instantaneous temperature change. |
Environmental Limitation Specifications:
M.O.L.E.® EV20 Profiler: •Maximum relative humidity 80% RH from 0°C to 31°C, decreasing linearly to 17% RH at 50°C. •Pollution Degree 2 (Normally only dry pollution, but with temporary conductivity caused by condensation) |
Battery Charger: •Temperature: (Operation) 0 to +45°C (32 to 113°F) •Temperature: (Non-Operation) -20 to +75°C (-4 to 167°F) •Humidity: (Operation) 20 to 90% |
NOTE:Tests have shown that because of the sensitive nature of the measurement and logic circuits, the following precautions must be observed: |
•Minimize exposure to ESD Events. If the M.O.L.E. Profiler or one of the thermocouples receives an 8kV electrostatic discharge during the data collection sequence, the M.O.L.E. Profiler may switch itself “OFF”. To retrieve the data, simply upload the data that was recorded prior to the event. If a new data collection sequence is started without uploading first, the existing data stored in the M.O.L.E. Profiler will be lost. |
•Keep strong electromagnetic fields away. The thermocouple wires serve as an antenna for electromagnetic radiation. If field strength of 3 volts per meter is present (usually due to close proximity of radio transmitters) while the M.O.L.E. Profiler is collecting data, the accuracy of the data may be compromised. This interference is compounded by using long thermocouple wires, as well as the effect of the thermocouple wires acting as a “tuned” antenna. |